Memmert HCP150 Humidity Chamber with 2 Stainless Steel Shelves 156L

Stock Code: 2776920
Manufacturer Part No: CHA4014
Order Now for 21 day delivery
Click Add to Quote for price and delivery
Quantity: - +

Applications for humidity chambers range from construction physics to corrosion testing and down to biological research. Ramp programming for temperature and humidity, active humidity control between 20 and 95 % rh as well as exact temperature control of up to +90 °C ensure a controlled, physiologically ideal environment for the simulation of real conditions.


ControlCOCKPIT: TwinDISPLAY with two high-resolution TFT colour touchdisplays


Setting temperature range 18 °C to 90 °C


Setting range active humidity control 20 to 95 % rh


Capacitive humidity sensor for measuring and displaying the relative humidity


Fully insolated stainless steel door and heated inner glass door


All-round heating on 6 sides


Ethernet interface for reading out protocol log


USB port on ControlCOCKPIT for use of USB stick


Multiple overtemperature protection/undertemperature alarm


Works calibration certificate at 60 °C / 75 % rh


Warranty is 3 years on parts, 1 year on Labour


TwinDISPLAY:


ControlCOCKPIT with two high-resolution TFT colour touchdisplays


Available parameters on the ControlCOCKPIT: Temperature (Celsius or Fahrenheit), fan speed, exhaust air flap position, program time, time zones, daylight-saving time


Two Pt100 sensors DIN class A in a 4-wire circuit for mutual monitoring, taking over functions in case of an error


HeatBALANCE function for application specific adjustment of heat output distribution (balance) between the upper and lower heating groups in an adjustment range between -50 % and +50 %


AtmoCONTROL software on a USB stick for programming, managing and transferring programs via Ethernet interface or USB port


Intuitive programming and archiving of ramps and program sequences


Synchronous visualisation of the created program sequence during programming


Application-specific repeat functions (loops) can be inserted within a temperature control program in any place


Simple creation of repeating weekly programs


ControlCOCKPIT with USB port for uploading programs, reading out protocol logs, activating the User-ID function


Displaying of already logged protocol data on the ControlCOCKPIT (max 10,000 values correspond to approx. 1 week)


Ethernet interface for reading out protocol log, uploading programs and for online logging


Multiple over temperature protection: Electronic temperature monitoring TWW/TWB (protection class 3.1 or 2) and mechanical temperature limiter TB (protection class 1) acc. to DIN 12880, AutoSAFETY automatically adjusts to the set value within a freely adjustable tolerance range. Setting individual


MIN / MAX values for over/under temperature alarm

Weight, kg90