Biochrom WPA Lightwave 3+ UV-Vis Spectrophotometer 3nm Bandwidth with USB Port In-Built Serial Print

Stock Code: 4082987
Manufacturer Part No: SPE1018
Order Now for 21 day delivery
Click Add to Quote for price and delivery
Quantity: - +

The Biochrom WPA Lightwave combines ease of use with flexibility, and the novel optics require no lid or moving parts and use a Xenon source for high energy performance with longer lamp lifetime. Compact footprint saves valuable bench space. This instrument has a 2 year warranty and the Xenon lamp has a 3 year warranty. The Lightwave 3+ is a higher resolution 3nm bandwidth version.


Powerful and flexible


The instrument includes a large wide view display and in-built software providing flash scan, fixed wavelength measurements, kinetics and concentration with comprehensive graphics capability plus the ability to store up to 90 methods. On peak confirmation is also a feature of the flexible software. Concentration may be measured using either a factor, single point calibration, multi-standard curves or alternatively there is a multi-wavelength mode where equations using absorbance values may be used for ratio calculations. Samples may be measured in 10, 20 or 40mm pathlength cells (glass, quartz or disposable).


Flexible data export options


All instruments have a USB port to download result data to a flash drive. Alternatively, results may be printed to an optional integrated graphical printer for a permanent record or exported via a USB cable connection to a suitable PC running Print Via Computer (PVC) software for advanced reporting and data storage or the optional wireless Bluetooth® accessory.


Split beam (RBC) UV/vis spectrophotometer covering 190 - 1100nm (choose from 3 or 5nm bandwidth versions)


Accepts 10, 20 and 40mm pathlength cuvettes


Compact, and excellent value for money compared to conventional systems


Wavelength scanning, kinetics and concentration functionality with full graphics display


Wavelength range: 190 to 1100 nm


Monochromator: flat grating


Wavelength calibration: automatic upon switch on


Beam height: 15 mm


Spectral bandwidth: 5nm or 3nm for '+' models


Wavelength accuracy: ±2nm


Wavelength reproducibility: ±1nm


Light source: xenon flash lamp


Detector: twin CMOS array


Photometric range: -0.300 to 2.500A, 0.3 to 199 %T


Photometric linearity: ±1.3 % or ±0.008 A whichever is greater at 546nm


Photometric reproducibility: ±0.002A to 0.5A at 546nm


Stray light: ﹤0.5 %T 340nm


Stability: ±0.01A/h at 340nm


Noise: ±0.005 peak to peak ±0.002 RMS


Dimensions: 260 x 390 x 100mm

Dims, w x d x h, mm260 x 390 x 100